Fluke PM-3394B Sander User Manual


 
3 - 44 USING THE COMBISCOPE INSTRUMENTS
3.8.4 Autoranging time base
The AUTO RANGE function of the Main Time Base (MTB) adjusts the time base
automatically, so that two to six waveform periods are displayed on the screen. If
a waveform doesn't contain enough information to calculate its period, the time
base is adjusted to acquire a minimum of two periods. One period of a signal is
determined by three successive crossings of the hysteresis band with the input
signal. The level of the hysteresis band can be set using the TRIGger:LEVel
command.
LIMITATION:
When operating with an acquisition length of 512 points, the maximum input
frequency is 25 MHz. For all other acquisition lengths, the maximum input
frequency is 50 MHz. When the input frequency is greater than the maximum alias
detection frequency, it is no longer possible to detect aliasing.
PROGRAM EXAMPLE:
CALL Send(0, 8, "INITiate:CONTinuous ON", 1)
Auto triggering
CALL Send(0, 8, "TRIGger:SOURce INTernal1", 1)
Sets CH1 trigger source
CALL Send(0, 8, "SENSe:SWEep:TIME:AUTO ON", 1)
Sets auto time base on
Figure 3.16 Definition of a signal period
PERI
O
D LEN
G
THPERI
O
D LEN
G
TH
ST7430
HYSTERESIS
TRIGGER
LEVEL
X2
X3X1