Sanyo MCO-19M(UVH) Welder User Manual


 
10
1. Outer door: The outer door is held to the frame with a magnetic seal. A door heater is installed in the
door panel. The door opening is reversible. Contact a Sanyo representative or agent to change the door
hinge from left to right or vice versa.
2. Inner door: Open when take the trays out or clean inside.
3. Gastight split door: Made of tempered glass, however avoid excessive impact on the glass.
4. Leveling feet: The leveling feet can be turned to adjust the height. Adjust the feet so that the Incubator
is level.
5. Trays:
Set trays at the same level as each gastight split door. Open a inner door when take the
tray out.
6. Tray supports: The tray supports can be removed by lifting the front side and pulling toward you.
7. Side supports: The right and left side supports can be removed for disinfection. Refer to page 45.
8. Duct: The duct for the path for circulating air. It is removable.
9. Fan cover: The fan cover serves as the inlet for circulating air. It is removable.
10. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an
autoclave.
11. Sample air outlet: The sample air outlet also functions as an internal gas outlet. Normally, cover this
outlet with the sample air outlet cap.
12. Sample air outlet cap: Always attach this cap except when using the sample air outlet.
13. Door switch: Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
2
and N
2
/O
2
when door is open. UV lamp is also deactivated by door opening*.
14. Humidifying pan: Fill the humidifying pan with sterile distilled water. Install the pan properly so that it
can be covered with the pan cover.
15. Humidifying pan cover: The cover prevents UV light entering the chamber. When filling the pan, lift
the front side and take out the pan. Refer to page 48 for details.
16. UV lamp*: This Sanyo UV lamp does not generate ozone. Never look directly at the UV light. For
replacement, contact a Sanyo representative or agent.
17. Water level sensor for humidifying pan: This sensor detects the water level in the humidifying pan.
Refer to page 44 for details.
18. Remote alarm terminals: Refer to page 13.
19. Access port: Place the silicon caps on the port both outside and inside when the port is not being
used.
20. Power switch: This is the main switch for the Incubator. It also functions as an overcurrent breaker.
21. Glow starter*: The glow started if for the UV lamp (model FG-7P)
22. Connecting port A/B for CO
2
gas pipe: When the optional MCO-21GC Automatic CO
2
Cylinder
Changeover System is installed, both ports A and B are available. If the MCO-21GC is not installed, only
port A is available. Refer to page 16 for gas cylinder connection. Ensure that the gas pressure is set at
0.03 MPaG (0.3 kgf/cm
2
G, 4.3 psiG). Refer to page 67 for automatic CO
2
cylinder changeover.