59
Example
Enter the value shown below if the displayed chamber CO
2
density is 5.0% but the actual CO
2
density is
4.5%.
●CO
2
density span: Enter 045.
Example
Enter the value shown below if the displayed chamber O
2
density is 5.0% but the actual O
2
density is
5.5%.
●O
2
density span: Enter 055.
4. Press the MENU Key to display the menu, select OK, and press the ENTER Key.
5. After the Select Tools Display is displayed, press MENU Key to display the menu, select Cancel, and
press the ENTER Key to return to the Top Display.
Note:
It is important to accurately measure the temperature inside the Incubator when calibrating the
temperature. Particularly, the temperature gauge used must have an accuracy of JIS 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting can not change by more than 1.0
o
C during calibration. If these values are
exceeded, an error buzzer will sound, the input data will be ignored, and the Incubator will return to the
display mode. If it is necessary to change the temperature by more than 1.0
o
C, perform calibration in
several stages over a period of time.
1
T e m p / C O 2 / O 2 C a l i b r a t i o n
3
P V
Volt
T e m p S p a n 0 . 0
o
C37 . 0
o
C 0 . 5°C
4
C O 2 Z e r o 3 . 9 1 V 5 0 0
5
C O 2 S p a n 4 . 5 % 5.0% 3. 16 V 7 0 0
O 2 Z er o 5 0 0
6
O 2 S p a n 0 . 0 % 5 . 0 % 5 8 m V 5 0 0
1
T e m p / C O 2 / O 2 C a l i b r a t i o n
3
P V
Volt
T e m p S p a n 0 . 0
o
C37 . 0
o
C 0 . 5°C
4
C O 2 Z e r o 3 . 9 1 V 5 0 0
5
C O 2 S p a n 0 . 0 % 5 . 0 % 3 . 1 6 V 7 0 0
O 2 Z er o 5 0 0
6
O 2 S p a n 5 . 5 % 5 . 0 % 5 8 m V 5 0 0