Sanyo MCO-19M(UVH) Welder User Manual


 
70
3. Input the CO
2
and O
2
density of standard gas to be used, press the MENU Key to display the menu,
select OK, and press the ENTER Key.
Ex. Standard gas density input
Standard gas density
(For example)
Input value of
CO
2
density
Input value of
O
2
density
Calibrate both of CO
2
and O
2
density CO
2
:5% / O
2
:5% 050 050
Calibrate only CO
2
density CO
2
:5% / O
2
:20% 050 200
Calibrate only O
2
density CO
2
:0% / O
2
:5% 000 050
Calibration method for CO
2
and O
2
gas density
1. Select Standard Gas Calibration from the Select Tools (2/2) Display, press the MENU Key to display
the menu, select OK, and press the ENTER Key.
2. The system check will start.
S e l e c t T o o l s ( 2 / 2 )
S t a n d a r d G a s S e t t i n g
S t a n d a r d G a s C a l i b r a t i o n
M
E N U
O K
S v c
C a n c e l
1
S t a n d a r d G a s C a l i b r a t i o n S t e p 1
3
S t d G a s C a l i b S y s t e m C h e c k
4
5
CO2 5. 5% O 2 4 . 5 %
STG 5. 0% STG 5 . 0 %
6
1
S t a n d a r d G a s S e t t i n g
3
S t d G a s C O 2 D e n s i t y 5 . 0 % ( 4 . 0 % - 2 1 . 0 % )
4
( N o Calibrat i o n : 0 . 0 % )
5
S t d G a s O 2 D e n s i t y 5 . 0 % ( 0 . 5 % - 1 8 . 0 % )
( 2 2 . 0 % - 8 1 . 0 % )
6
( N o Calibration:18.1 % - 2 1 . 9%)
M
E N U
O K
C a n c e l