47
12. Clean all the attachments with a diluted neutral detergent, and then rinse them thoroughly with
distilled water.
13. Wipe the trays, the inner attachments such as the chamber circulation fan, and the chamber sides
with sterilizing alcohol. Be careful not to leave any residual alcohol.
14. Wipe the water level sensor with sterilizing alcohol. Be careful not to leave any residual alcohol. When
cleaning the sensor, take care not to apply excessive force to the lead wires.
15. To reinstall all the attachments, perform the procedure in reverse order from step 14.
Note:
• When installing the fan, insert it securely on the motor shaft (See Fig.11). Lightly turn the fan manually
to make sure that it does not strike the rear panel. Improper insertion may cause poor performance.
• As shown in Fig. 12, set the tray with only the front edge bent down. If the tray is set in the wrong
direction, it may not be level and may become unstable.
Fig. 11
Fig. 12
Bent down
Fan
Spring