Sanyo MCO-19M(UVH) Welder User Manual


 
25
BASIC PARAMETERS
Setting the chamber temperature, CO
2
and O
2
density
The setting procedure for the chamber temperature, CO
2
and O
2
density are given below. (Default
settings: Chamber temperature: 37
o
C, CO
2
density: 0%, O
2
density: 20%)
The Incubator automatically starts operation using these settings after the power is turned ON.
1. From the Top Display, press the MENU Key to access the menu. Select Set and press the ENTER
Key.
2. The Stand-By Setting Display will be displayed. Set the parameters.
The setting ranges of the parameters are as follows:
Temperature: 0 to 50
o
C. To 37.0
o
C, enter 370.
CO
2
density: 0 to 20%. To set 5.0%, enter 050.
O
2
density: 1 to 18 and 22 to 80%. To set 5.0%, enter 050.
3. Press the MENU Key to display the menu after setting the parameters. Select OK and press the
ENTER Key. The settings will be saved.
Note:
When starting the Incubator for the first time or after not using it for an extended period of time, allow at
least 4 hours for the chamber temperature, humidity, CO
2
sensor and O
2
sensor to become stable after
setting the desired chamber temperature at a 0% CO
2
density and a 20% O
2
density. Then change the
setting to the desired CO
2
and O
2
density.
.
1
T e m p 3 7 . 0
o
C C O 2 5 . 0 %
O 2 5 . 0 %
2
3
4
5
S t a t u s
6
O K
7
.
.
.
M E N U
S e t
L o g
T o o l s
1
S t a n d - by S e t t ing
2
T e m p e r a t u r e
3 7 . 0
o
C ( 0. 0
o
C - 5 0 . 0
o
C)
3
C O 2 D e n s i t y 0 . 0 %
( 0 . 0 % - 2 0 . 0 % )
4
O 2 D e n s i t y 2 0 . 0 %
( 1 . 0 % - 1 8 . 0 % )
5
( 2 2 . 0 % - 8 0 . 0 % )
K e y L o ck 0 (0.Unlock 1 . L o c k)
6
H i g h L i m i t 5 2 . 0
o
C
M E N U
O K
C a n c e l