Preventive Maintenance
Flame photometric detector
194
Likewise, damage to the PMT window cannot be tolerated; if necessary,
replace the PMT or call Hewlett•Packardsupport.
1. Remove four screws to remove the PMT adapter flange. Remove the
adapter carefully; a quartz window is exposed and may fall out. The
window is cleaned in a manner similar to the filter.
2. Remove four more screws to remove the stainless steel coupling.
Remove the coupling carefully; the remaining quartz window may fall
out. Clean the window in a manner similar to the filter.
Note placement and types of seals found on disassembled parts. Seals
should be replaced on reassembly. (See Figure 8•21for seal part
numbers.)
Inspect the windows; chips missing, scratches, and/or cracks in the light
path scatter light, reducing detector sensitivity. Replace window(s) if
necessary (first window•Part No. 19256•80030;second window•Part
No. 19256•80060.)
Reassemble parts in reverse order, making sure all seals are of the proper
type and in their respective proper locations. Tighten screws firmly to
ensure gas•tightand light•tightseals. If the filter has a silvered side, it
should face the flame (indicator arrows (>) on edge of filter should point
toward the PMT).