For analytical information from a detector, proper settings for
RANGE 2 ()
and
ATTN 2 ()
are determined such that peaks of interest are on scale
at the integrator or chart recorder: peaks of interest must neither flat
top by exceeding the allowed maximum output level, nor be too small to
be measured. Table 5•3lists maximum detector output producing +1 volt
at the +1V output for each
RANGE 2 ()
setpoint value.
FID & NPD (pA)
TCD
(mV, High Gain)
TCD
(mV, Low Gain)
ECD (kHz)
Maximum Detector Signal
Producing +1 V Output
0 1.0 10
3
25 800 10
1 2.0
10
3
50 20
2 4.0 10
3
40
3 8.0
10
3
80
4 1.6 10
4
160
5 3.2
10
4
320
6 6.4 10
4
7 1.3 10
5
8 2.6 10
5
9 5.1 10
5
10 1.0 10
6
11 2.0 10
6
12 4.1 10
6
13 8.2 10
6
RANGE 2 ()
Table 5-3. Detector Output Producing a 1 V Signal