A SERVICE OF

logo

Test Sample Chromatograms
Test sample chromatograms
227
Detector Type ECD(or ECDw/MUG)
Temp 300 DEGREES C
Inlet Type PACKED
(OR PURGED
PACKED).
Temp 200
DEGREES C
Operating Mode N/A
Purge Time On N/A min
Purge Time Off N/A
min
Oven Temp Isothermal
Init Temp 160
DEGREES C
Init Time N/A
min
Ramp
Rate 0
Fin Temp
Fin Time
Range 2
COLUMN:
Part No. 19095S(#100)
Dimensions 530 MID; 5 m
Sta Phase Methyl
Silicone
START
.88
.12
2.
5
STOP
FLOW RATES
Carrier (N2) 30 +/• 1 ml/min
Hydrogen N/A ml/min
Air N/A ml/min
Makeup (N2) 30 +/• 1 ml/min
Split Vent N/A ml/min
Septum Purge 1•2 ml/min
SAMPLE:
Type ECD Sample
Inj Volume 1
l
Part No. 18713•60040
Composition 33 pg/
l(0.033
ppm(W/V)) each: and
aldrin in isooctane
LINDANE
ALDRIN
HP 5890 Test Sample Operating Conditions
Electron Capture Detector (ECD)
Figure 10-3.