Agilent Technologies 6843A Welding System User Manual


 
3 - Language Dictionary
112
TRIGger:SEQuence3:SOURce
TRIGger:ACQuire:SOURce
These commands select the trigger source for a triggered measurement sequence as follows:
BUS
GPIB device, *TRG, or <GET> (Group Execute Trigger)
EXTernal
ac source’s backplane Trigger In BNC
TTLTrg
the signal driving the Trigger Out BNC
Command Syntax
TRIGger:SEQuence3:SOURce <CRD>
TRIGger:ACQuire:SOURce <CRD>
Parameters
BUS | EXTernal | TTLTrg
*RST Value
BUS
Examples
TRIG:ACQ:SOUR BUS TRIG:SEQ3:SOUR EXT
Query Syntax
TRIGger:SEQuence3:SOURce?
TRIGger:ACQuire:SOURce?
Returned Parameters
<CRD>
Related Commands
ABOR TRIG TRIG:DEL TRIG:SYNC
TRIG:SYNC:PHAS INIT INIT:CONT *TRG WAI
TRIGger:SEQuence1:DEFine
TRIGger:SEQuence2:DEFine
TRIGger:SEQuence3:DEFine
These commands define the names that are aliased to trigger sequences 1, 2 and 3. The command
accepts only TRANsient for sequence 1, SYNChronous for sequence 2, and ACQuire for sequence 3 as
predefined names. The query allows the user to query the instrument names aliased to sequences 1, 2,
and 3.
Command Syntax
TRIGger:SEQuence1:DEFine TRANsient
TRIGger:SEQuence2:DEFine SYNChronous
TRIGger:SEQuence3:DEFine ACQuire
Parameters
TRANsient, SYNChronous, ACQuire
*RST Value
BUS
Examples
TRIG:SEQ1:DEF TRAN TRIG:SEQ3:DEF ACQ
Query Syntax
TRIGger:SEQuence1:DEFine?
TRIGger:SEQuence2:DEFine?
TRIGger:SEQuence3:DEFine?
Returned Parameters
<CRD>