VOLTage:OFFSet:MODE
Agilent 6811B, 6812B, 6813B, Only
This command determines how the dc offset voltage is controlled during a triggered output transient. The
choices are:
FIXed
The offset is unaffected by a triggered output transient.
STEP
The offset is programmed to the value set by VOLTage:OFFSet:TRIGgered when a
triggered transient occurs.
PULSe
The offset is changed to the value set by VOLTage:OFFSet:TRIGgered for a duration
determined by the pulse commands.
LIST
The offset is controlled by the voltage list when a triggered transient occurs.
Command Syntax
[SOURce:]VOLTage:OFFSet:MODE <mode>
Parameters
FIXed | STEP | PULSe | LIST
*RST Value
FIXed
Examples
VOLT:OFFS:MODE FIX VOLT:OFFS:MODE:LIST
Query Syntax
[SOURce:]VOLTage:OFFSet:MODE?
Returned Parameters
<CRD>
Related Commands
VOLT:OFFS VOLT:OFFS:TRIG
VOLTage:OFFSet:TRIGgered
Agilent 6811B, 6812B, 6813B, Only
This command selects the dc offset that the output waveform will be set to during a triggered step or pulse
transient.
The maximum peak voltage that the ac source can output is 425 V peak. This includes any combination of
voltage, voltage offset, and function shape values. Therefore, the maximum value that can be
programmed depends on the peak-to-rms ratio of the selected waveform. For a sinewave, the maximum
voltage that can be programmed is 300 V rms.
NOTE: The OUTPut:COUPling must be set to DC to get non-zero dc output.
Command Syntax
[SOURce:]VOLTage:OFFSet:TRIGgered <NRf+>
Parameters
–425 to +425 | MAXimum | MINimum
Unit
V
(dc voltage)
*RST Value
0
Examples
VOLT:OFFS:TRIG 50 VOLT:OFFS:TRIG INF
Query Syntax
[SOURce:]VOLTage:OFFSet:TRIGgered?
Returned Parameters
<NR3>
Related Commands
VOLT:OFFS:MODE OUTP:COUP