6. While holding the sensor in position, tighten the hex
nut of the male connector to firmly secure the sensor
in place. When the hex nut is tightened, the Teflon
ferrule inside the compression fitting clamps the sen-
sor tube.
NOTE
A stainless steel ferrule is available if the
Teflon ferrule does not inadequately grip.
When using the metallic ferrule, care must be
taken to avoid over tightening and damaging
the sensor tube. If the male connector leaks
during insertion or retraction, replace the O-
ring in the male connector.
2.2.2 INSTALLATION WITHOUT A BALL VALVE. The
Model 396R Sensor may be installed through a weldalet
or pipe tee or “Y” when used with a process connector
(PN 23166-00 or 23166-01). The sensor should be
installed within 80° of vertical, with the electrode facing
down.