Fluke 187 Welder User Manual


 
Model 187 & 189
Users Manual
3-8
Keep the following in mind when measuring resistance:
Because the meters test current flows through all
possible paths between the probe tips, the measured
value of a resistor in a circuit is often different from
the resistors rated value.
The test leads can add 0.1 to 0.2 of error to
resistance measurements. To test the leads, touch
the probe tips together and read the resistance of the
leads. If necessary, you can press D to
automatically subtract this value.
The resistance function can produce enough voltage to
forward-bias silicon diode or transistor junctions, causing
them to conduct. To avoid this, do not use the 30 M or
500 M ranges for in-circuit resistance measurements.
Testing for Continuity
Caution
To avoid possible damage to the meter or to
the equipment under test, disconnect circuit
power and discharge all high-voltage
capacitors before testing for continuity.
Continuity is the presence of a complete path for current
flow. The continuity test features a beeper that sounds if a
circuit is complete. The beeper allows you to perform
quick continuity tests without having to watch the display.
The continuity function detects intermittent opens and
shorts lasting as little as 1 millisecond (0.001 second).
These brief contacts cause the meter to emit a short beep.
To select continuity, turn the rotary switch to resistance
position, then press the blue button once. The continuity
symbol (~) appears in the display. Continuity uses
manual ranging only; autoranging is not available. Refer to
Figure 3-6 for continuity testing setup instructions.