Tektronix TDS430A Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–109
DESE?
might return the string :DESE 186, showing that the DESER contains the binary
value 10111010.
DIAg:RESUlt:FLAg? (Query Only)
Returns the pass/fail status from the last diagnostic test sequence execution. The
DIAg:RESUlt:LOG? query can be used to determine which test(s) has failed.
Calibration and Diagnostic
DIAg:RESUlt:LOG?
DIAg:RESUlt:FLAg?
:
RESUlt
DIAg :
?
FLAg
PASS indicating that all of the selected diagnostic tests have passed.
FAIl indicating that at least one of the selected diagnostic tests has failed.
DIAG:RESULT:FLAG?
returns either PASS or FAIl.
DIAg:RESUlt:LOG? (Query Only)
Returns the internal results log from the last diagnostic test sequence execution.
The list contains all modules and module interfaces that were tested along with
the pass/fail status of each.
Calibration and Diagnostic
DIAg:RESUlt:FLAg?
DIAg:RESUlt:LOG?
Group
Related Commands
Syntax
Returns
Examples
Group
Related Commands
Syntax