Tektronix TDS430A Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–131
VARpersist leaves acquired data points on the display for a period of time
specified by DISplay:PERSistence.
VECtors connects adjacent data points. New points immediately replace old
ones.
INTENSIFied causes the display to show acquired (non-interpolated) samples
with brighter or different colored dots than the rest of the waveform.
DISPLAY:STYLE INFPERSIST
sets the display to indefinitely accumulate data points on the screen.
DISPLAY:STYLE?
might return DOTS indicating that the display shows individual waveform data
points with no history.
DISplay:TRIGBar
Controls the display of the trigger level indicator bar/s. The indicator bar shows
where the trigger will occur, in voltage.
The digitizing oscilloscope will only display the bar if the trigger source is also
displayed. If both a main and a delayed trigger are displayed, then two bars will
appear. One will accompany each source. If a logic trigger is selected, then
multiple bars may appear. If a runt pulse trigger is selected, then two bars may
appear. One will show the upper threshold and one the lower threshold.
Display
DISplay:TRIGBar { OFF | SHORt | LONG }
DISplay:TRIGBar?
TRIGBar
SHORt
DISplay
:
<Space>
?
OFF
LONG
OFF removes the trigger indicator bar from the display.
SHORt displays, as the indicator, a short arrow at the right side of the graticule for
each displayed trigger signal.
Examples
Group
Syntax
Arguments