Tektronix TDS430A Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–313
TRIGger
:
:
PULse
MAIn
SLEWRate
:
:
?
<Space>
<NR3>
DELTATime
<NR3> is the delta time in seconds.
TRIGGER:MAIN:PULSE:SLEWRATE:DELTATIME 15E-6
sets the slew rate trigger’s delta time to 15 ms.
TRIGger:MAIn:PULse:SLEWRate:POLarity
TDS 500C, 600B & 700C Only
Sets or queries the polarity for the main pulse slew rate trigger. This command is
equivalent to selecting Polarity in the Trigger menu with Type set to SlewRate.
Trigger
TRIGger:MAIn:PULse:SLEWRate:POLarity { EITher | NEGAtive |
POSITIVe }
TRIGger:MAIn:PULse:SLEWRate:POLarity?
<Space>
EITher
?
POSITIVe
NEGAtive
TRIGger :
MAIn
POLarity
:
PULse
:
:
SLEWRate
NEGAtive indicates that a pulse edge must traverse from the upper (most
positive) to lower (most negative) level for slew rate triggering to occur.
POSITIVe indicates that a pulse edge must traverse from the lower (most
negative) to higher (most positive) level for slew rate triggering to occur.
EITher indicates either NEGAtive or POSITIVe polarity.
TRIGGER:MAIN:PULSE:SLEWRATE:POLARITY EITHER
specifies that the polarity of the slew rate can be either positive or negative.
Arguments
Examples
Group
Syntax
Arguments
Examples