Agilent Technologies 6812B Welding System User Manual


 
3 - Language Dictionary
44
MEASure:CURRent:CREStfactor?
FETCh:CURRent:CREStfactor?
Phase Selectable
These queries return the output current crest factor. This is the ratio of peak output current to rms output
current.
Query Syntax
MEASure:[SCALar]:CURRent:CREStfactor?
FETCh:[SCALar]:CURRent:CRESfactor?
Parameters
None
Examples
MEAS:CURR:CRES? FETC:CURR:CRES?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:ACDC? MEAS:CURR:AMPL:MAX?
MEASure:CURRent:HARMonic?
FETCh:CURRent:HARMonic?
Phase Selectable
These queries return the rms amplitude of the Nth harmonic of output current.
The parameter is the desired harmonic number. Queries sent with a value of 0 return the dc component. A
value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the
fundamental measurement bandwidth of the measurement system, which is 12.6kHz. Thus the maximum
harmonic that can be measured is dependent on the output frequency. Any harmonics that represent
frequencies greater than 12.6kHz are returned as 0.
Query Syntax
MEASure:[SCALar]:CURRent:HARMonic[:AMPLitude]? <NRf>
FETCh:[SCALar]:CURRent:HARMonic[:AMPLitude]? <NRf>
Parameters
0 to 50
Examples
MEAS:CURR:HARM? 3 FETC:CURR:HARM? 1
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:HARM:PHAS? MEAS:CURR:HARM:THD?