Agilent Technologies 6812B Welding System User Manual


 
Language Dictionary - 3
59
OUTPut:RI:MODE
This command selects the mode of operation of the Remote Inhibit protection. The following modes can
be selected:
LATChing
A TTL low at the RI input latches the output in the protection shutdown state, which can
only be cleared by OUTPut:PROTection:CLEar.
LIVE
The output state follows the state of the RI input. A TTL low at the RI input turns the output
off; a TTL high turns the output on.
OFF
The instrument ignores the RI input.
Command Syntax
OUTPut:RI:MODE <mode>
Parameters
LATChing | LIVE | OFF
*RST Value
LATChing
Examples
OUTP:RI:MODE LIVE
Query Syntax
OUTPut:RI:MODE?
Returned Parameters
<CRD>
Related Commands
OUTP:PROT:CLE
OUTPut:TTLTrg
This command enables or disables the ac source Trigger Out signal, which is available at a BNC
connector on the rear of the instrument.
Command Syntax
OUTPut:TTLTrg[:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
OUTP:TTLT 1 OUTP:TTLT ON
Query Syntax
OUTPut:TTLTrg[:STATe]?
Returned Parameters
0 | 1
Related Commands
OUTP:TTLT:SOUR
OUTPut:TTLTrg:SOURce
This command selects the signal source for the Trig Out signal as follows:
BOT Beginning of transient output
EOT End of transient output
LIST Specified by the TTLTrg list
When an event becomes true at the selected TTLTrg source, a pulse is sent to the BNC connector on the
rear of the ac source.
Command Syntax
OUTPut:TTLTrg:SOURce <source>
Parameters
BOT | EOT | LIST
*RST Value
BOT
Examples
OUTP:TTLT:SOUR LIST
Query Syntax
OUTPut:TTLTrg:SOURce?
Returned Parameters
<CRD>
Related Commands
OUTP:TTLT