The Tester has two types of configurations: standard and power-up. The
standard configuration, which is programmed into the Tester and can always
be retrieved, includes the following settings:
FCN V Volts Function selected.
DISP WAVE Waveform Screen Mode selected.
THD %RMS Total Harmonic Distortion computed as percentage of
the total RMS value.
CLAMP 80I-500S The 80i-500s Current Probe is selected.
OTHER specifies any current probe other than the 80i-500s. The Tester
has been calibrated to provide either compensation for the 80i-500s or a
flat response for other probes.
REF V Voltage Phase Reference selected.
For Model 41B,
KBAUD 9.6 is the standard configuration for Print or Send,
and
EPSON is the standard configuration. The Tester always selects
autoranging at power-up.
The power-up configuration, which is initially the same as the standard
configuration, can be changed by making entries from the configuration
screen. The Tester sets up to the power-up configuration each time it is turned
on. You can change the power-up configuration using the following
procedure:
1. Press
while pressing to turn the Tester on. The configuration
screen appears. (See Figure 17.)
2. Press
and to highlight different items on the same line. Press
to select the currently highlighted item and proceed to the next
line.