Use, Maintenance, and Care of the Devices
Using the Sliding Load
Using the Sliding Load
When performing a sliding load calibration, it is recommended that the sliding ring be set
at the marked positions (rings) along the sliding load body. Using the set marks ensures
that a broad distribution of phase angles is selected, thereby optimizing the calibration.
The set marks function as detents so that the internal center of the sliding ring can mate
with them. Because of this, the set mark being used cannot be seen but is felt as the sliding
ring is moved from mark to mark during a calibration. Moving the sliding ring with only
the index fingers of both hands will increase your ability to detect the sliding ring detent at
each position.
To perform a sliding load calibration, refer to your network analyzers user’s documentation
for instructions.