Standard Definitions
Standard Class Assignments
Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error
models used in the measurement calibration. A class or group of classes corresponds to the
systematic errors to be removed from the measured network analyzer response. Table A-1
through A-3 list the classes of the devices in the kit for various network analyzers. This
information resides on the calibration definitions disk included in the kit.
Table A-1 Standard Class Assignments for the 8510 Network Analyzer
Disk File Name: CK_NTYPB2 Calibration Kit Label: TYPE N B.2
File Number: * FILE 1
Class A B C D E F G Standard Class Label
S
11
A24 Open
S
11
B13 Short
S
11
C 9 10 12 Loads
S
22
A24 Open
S
22
B13 Short
S
22
C 9 10 12 Loads
Forward transmission 11 Thru
Reverse transmission 11 Thru
Forward match 11 Thru
Reverse match 11 Thru
Forward isolation
a
a. The forward isolation standard is also used for the isolation part of the response and isolation
calibration.
9 Isol’n Std
Reverse isolation 9 Isol’n Std
Frequency response 132411 Response
TRL thru 14 Undefined
TRL reflect 1 Undefined
TRL line 15 Undefined
Adapter 13 14 Adapter
TRL Option
Cal Z
0
: _____ System Z
0
__X__ Line Z
0
Set ref: __X__ Thru _____ Reflect
Lowband frequency: 2.0 GHz