A-14 85054B
Standard Definitions
Nominal Standard Definitions
Table A-9 Standard Definitions for the PNA Series Network Analyzer
System Z
0
a
= 50.0
Ω
Calibration Kit Label:
N 50Ω
Standard
b
C0
×
10
−18
F
C1
×
10
−30
F/Hz
C2
×
10
−39
F/Hz
2
C3
×
10
−48
F/Hz
3
Fixed or sliding
Offset
Frequency
in GHz
c
Coax or Waveguide
Standard Label
Number
Type
L0
×
10
−12
H
L1
×
10
−24
H/Hz
L2
×
10
−33
H/Hz
2
L3
×
10
−45
H/Hz
3
Delay in ps
Z
0
Ω
Loss in G
Ω
/s
Min
Max
1Short
−
0.1315
606.21 −
68.405
2.0206
27.990 50 1.3651 0 999 Coax
Short (m)
d
2Open104.13
−
1943.4
144.62 2.2258
22.905 50 0.93 0 999 Coax
Open (m)
d
3Broadband
load
Fixed 0 50 0 999 Coax Broadband
load
4 Thru Thru 0 50 0 999 Coax Thru
5 Sliding
load
Sliding 0 50 0 999 Coax Sliding
load
6Lowband
load
Fixed 0 50 0 999 Coax Lowband
load
7 Short 0.7536 459.88
−
52.429 1.5846 63.078 50 1.1273 0 999 Coax
Short (f)
d
8 Open 89.939
2536.8 −
264.99
13.4
57.998 50 0.93 0 999 Coax
Open (f)
d
a. Ensure system Z
0
of network analyzer is set to this value.
b. Open, short, load, delay/thru, or arbitrary impedance.
c. For waveguide, the lower frequency is the same as F
CO.
d. Standard labels which specify ex, (m) or (f), refer to the sex of the test port connector.