Agilent Technologies E4400-90326 Welding System User Manual


 
Quick Start Guide 1-9
ESG Family Signal Generators Getting Started
Saving and Recalling an Instrument State
Saving and Recalling an Instrument State
Using this procedure, you will learn how to save instrument settings to a memory register and
to recall the settings.
Creating an Instrument State
1. Press Preset. This sets the signal generator to its factory-defined instrument
state.
2. Press Frequency. Frequency becomes the active function and the preset value for
frequency is displayed in the active entry area.
3. Enter 800 using the numeric
keypad and press the MHz
softkey.
The new carrier frequency (800.000 000 00 MHz) is
shown in both the FREQUENCY area and the active entry area
of the display.
4. Press Amplitude. Amplitude becomes the active function, and the preset value for
amplitude is displayed in the active entry area.
5. Enter 0 using the numeric
keypad and press the dBm
softkey.
The new power level (0.00 dBm) is shown in both the
AMPLITUDE area and the active entry area of the display.
6. Press AM followed by the
AM Off On softkey.
AM toggles from Off to On. Notice also that the AM display
annunciator is turned on, indicating that you have enabled
amplitude modulation. You have now created an example
instrument state that you will save and recall.
qsg.book Page 9 Friday, September 22, 2000 3:13 PM