Emerson MSP400RH Laser Level User Manual


 
45
IP2046/IM
Nov 2006
base units m ft in
SETUP DUTY P010 Bottom Reference as base units 11.0 36.0 432.0
P011 Tank Shape - Linear Linear Linear
P013 PV Scale Factor - 1.0 1.0 1.0
P014 Profile Height as base units 1.0 1.0 1.0
NLP CURVE P030 Profile Point 1 % 10 10 10
P031 Profile Point 2 % 20 20 20
P032 Profile Point 3 % 30 30 30
P033 Profile Point 4 % 40 40 40
P034 Profile Point 5 % 50 50 50
P035 Profile Point 6 % 60 60 60
P036 Profile Point 7 % 70 70 70
P037 Profile Point 8 % 80 80 80
P038 Profile Point 9 % 90 90 90
P039 Profile Point 10 % 100 100 100
> P000 Message -
> P001 Tag -
> P002 Description -
PV CALC P012 Primary Variable Units (PV Units) - m ft in
OUTPUT CURRENT P015 Upper range value L000 10.55 34.5 414.0
P016 Lower range value L000 0.0 0.0 0.0
P020 Damping sec 3.0 3.0 3.0
L000 Range Value Units -
RELAYS RELAY 1 P070 Relay 1 mode -
P071 Relay 1 PV ON Point as PV units 0.0 0.0 0.0
P072 Relay 1 PV OFF Point as PV units 0.0 0.0 0.0
RELAY 2 P073 Relay 2 mode -
P074 Relay 2 PV ON Point as PV units 0.0 0.0 0.0
P075 Relay 2 PV OFF Point as PV units 0.0 0.0 0.0
ENGINEERING P021 LE Delay sec 900 900 900
P022 LE Action - Hold Hold Hold
P023 Blanking base units 0.45 1.5 18.0
P024 Speed of Sound base units/s 331.8 1088.6 13063
P025 Temperature C or F Auto Auto Auto
P026 Threshold % Auto Auto Auto
ADVANCED P041 Pulse Repetition sec 1.0 1.0 1.0
P042 Echoes Needed - 4 4 4
P043 Threshold 1 Time ms 2.0 2.0 2.0
P044 Target Pulses - Auto Auto Auto
P045 Target Frequency kHz Auto Auto Auto
P049 Spike Rejection - 0 0 0
SYSTEM AUTO CYCLE -
LOAD DEFAULTS -
L200 Base Units -
FIXED P004 Final Assembly Number -
P005 Serial Number -
P970 Front face material - Kynar Kynar Kynar
HART D949 Model Code - 53 53 53
D950 HART Device Code - 46 46 46
D951 Comms Address -
D952 Hardware Revision -
D953 Software Version -
D960 Manufacturer's Code -
D961 Unique ID -
D962 Universal Cmd Rev - 5 5 5
D963 Transmitter Spec. Cmd Rev -
D964 Response Preamble - 5 5 5
D965 Transmitter Flags -
MONITOR READINGS VARIABLES D900 Primary Variable base units
D901 Level (SV) base units
D902 Range (TV) base units
D903 Transducer Temperature C or F
CURRENT D906 Current output mA
D905 % Current Output %
> D908 Relay Status -
USER DEFAULTS
MESSAGE
MSP400 XMTR
MSP400
Unchanged (ex-factory = 0)
as applicable
as applicable
Solartron Mobrey
as applicable
Fault/Setpoint
Setpoint
MSP400RH
as applicable
as applicable
as applicable
DIAGNOSTICS D910 Target Range / Distance to Target base units
D911 Echo Size %
D912 Echo Success Rate %
D913 Target Echoes -
D914 Speed of Sound base units/s
D915 Transducer Temperature C or F
D916 Transducer Frequency kHz
HISTORY P003 Date of Change dmy 01/01/02 01/01/02 01/01/02
P046 Maximum Temperature °C 50 50 50
P047 Minimum Temperature °C -10 -10 -10
D 3.0 Table of MSP400 Parameters