Agilent Technologies E4443A Saw User Manual


 
Chapter 3 205
Instrument Functions: M - O
MODE
Instrument Functions: M - O
SCPI Status Bits/
OPC Dependencies: If you are using the status bits and the analyzer mode is changed, the
status bits should be read, and any errors resolved, prior to switching
modes. Error conditions that exist prior to switching modes cannot be
detected using the condition registers after the mode change. This is true
unless they recur after the mode change, although transitions of these
conditions can be detected using the event registers.
ChangingmodesresetsallSCPIstatusregistersandmaskregistersto
their power-on defaults. Hence, any event or condition register masks
must be re-established after a mode change. Also note that the power up
status bit is set by any mode change, since that is the default state after
power up.
Factory Preset: 1 (Spectrum Analysis)
If
Preset Type Mode is selected, then the analyzer settings are preset but it
stays in that selected mode.
Remote Command:
:INSTrument:NSELect <integer>
:INSTrument:NSELect?
Remote Command Notes: Enter one of the following integers in the command to set the
analyzer mode.
Example: INST:NSEL 4
Mode NSELect
Number
Mode Keyword
Basic 8 BASIC
cdmaOne 4 CDMA
CDMA1xEV-DO 15 CDMA1XEV
cdma2000 10 CDMA2K
EDGE with GSM 13 EDGEGSM
89600 VSA Link Software 231 LINK
NADC 5 NADC
Noise Figure 219 NFIGURE
PDC 6 PDC
Phase Noise 14 PNOISE
Spectrum Analysis 1 SA
W-CDMA for 3GPP 9 WCDMA