Agilent Technologies E4443A Saw User Manual


 
Chapter 5 401
One-Button Measurement Functions
Meas Setup (Channel Power—CHP)
One-Button Measurement Functions
Factory Preset: 2 MHz, or as defined by the selected radio standard.
Terminators: GHz, MHz, kHz, Hz
Default Terminator: Hz
Knob Increment: 1%
Step Key
Increment: 1,1.5,2,3,5,7.5,10...
Range: 100 Hz to maximum span supported by your analyzer.
Remote Command:
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
Example: CHP:BAND:INT 1 MHz
CHP:BAND:INT?
5.14.4 Chan Pwr Span
Press Chan Pwr Span to set the analyzer span for the channel power measurement. When the
RRC Filter is set to On, the lower limit for the span is .
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 3 MHz, or as defined by the selected radio standard.
Default Terminator: Hz
Knob Increment: 1, 1.5, 2, 3, 5, 7.5, 10...
Step Key
Increment: 1%
Range: Current integration bandwidth to 10 times the integration bandwidth or
span of your analyzer.
Remote Command:
[:SENSe]:CHPower:FREQuency:SPAN <freq>
[:SENSe]:CHPower:FREQuency:SPAN?
Example: CHP:FREQ:SPAN 2 MHz
CHP:FREQ:SPAN?
1alpha+()Integration BW×