Agilent Technologies E4443A Saw User Manual


 
488 Chapter5
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
5.27.3.1 Range
Selects a range and updates the values on the other
Range Table keys so that they reflect the
settings for the selected range. If
Range is set to On it is used as part of the measurement;
when set to
Off it is excluded.
Key Path:
Meas Setup, Range Table
Factory Preset: See “Range Table” on page 486.
Range: Off|On
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted. For example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values remain
as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe OFF|ON|0|1
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:STATe?
Example: SPUR:STAT OFF,ON,ON,OFF,ON,ON
listsvalues1through6
5.27.3.2 Start Freq
Used to set the start frequency of the analyzer.
Key Path:
Meas Setup, Range Table
Factory Preset: See “Range Table” on page 486.
Knob Increment: 1%
Step Key
Increment: 1, 1.5, 2, 3, 5, 7.5, 10,...
Range: Frequency range of your analyzer.
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:FREQuency:STARt <integer>
The query for this parameter will always return 20 values.