Agilent Technologies E4443A Saw User Manual


 
466 Chapter5
One-Button Measurement Functions
Meas Setup (Spectrum Emissions Mask—SEM)
One-Button Measurement Functions
values)
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime?
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:SWEeptime:AUTO?
Remote Command Notes: Comma separated list of up to 5 values. n = 1 for BTS and n = 2
for MS. Default is BTS.
Example: SEM:OFFS:LIST:SWE:AUTO 4 ms
SEM:OFFS:LIST:SWE:AUTO?
SEM:OFFS:LIST:SWE 4 ms
SEM:OFFS:LIST:SWE?
5.22.4.5 Res BW
Specifies the resolution bandwidth used in measuring the offset pair. When set to Auto, the
Res BW from the default tables for
“Offset/Limits” on page 462 are used. When set to Man,
the range of settings is the range of the available Res BW of the analyzer, except the
maximum is further limited to not exceed (Stop Freq - Start Freq).
Key Path:
Meas Setup, Offset/Limits
State Saved: Saved in instrument state.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for
“Offset/Limits” on page 462 for this value.
Default Terminator: Hz
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]
<freq>...[<freq>] (uptofivevalues)
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]?
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO
OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth[:RESolution]:AUTO?
Remote Command Notes: Comma separated list of up to5 values. n = 1 for BTS and n = 2
for MS. Default is BTS (1).
You may only enter valid Res BW filter frequencies. Refer to
“Res BW” on
page 83
.