Agilent Technologies E4443A Saw User Manual


 
Chapter 4 255
Instrument Functions: P - Z
System
Instrument Functions: P - Z
restored by System, Restore Sys Defaults.
SCPI Status Bits/
OPC Dependencies: None
Factory
Default: Not affected by “
Restore Sys Defaults.”
Terminators: Unitless
Range: Year (YYYY): 1970 to 2029
Month (MM):01 to 12
Day: 01 to 30 or 31 (depending on the month)
Remote Command:
:SYSTem:DATE <year>,<month>,<day>
:SYSTem:DATE?
Example: SYST:DATE 2000,12,24 Sets the date to December 24, 2000
4.8.4 Alignments
Displays functions that control the automatic alignment of the instrument and load
default values for the alignment system.
NOTE Most CALibration commands execute in the background, permitting other
SCPI commands to be processed concurrently. If a measurement command is
sent right after a CALibration command, there can be interaction between
background alignments and the measurement. The *WAI command should be
issued after any CALibration command and before the measurement
command. Note that sending the query form of a CAL? command will
automatically hold off any following commands until the query value is
returned.
Bit 0 (CALibration) must be set to 1 in the STATus:OPERation:ENABle
register to ensure that the *WAI command waits for calibration to complete.
This is the factory preset and *RST setting.
Key Path:
System
Remote Command:
There is no remote command for this key.
4.8.4.1 Auto Align
Allows you to turn the instrument automatic alignment
On or Off. or select Alert to be