Fluke PM-3380B Sander User Manual


 
4 - 88 COMMAND REFERENCE
SENSe:VOLTage<n>[:DC]:RANGe:PTPeak
Syntax:
SENSe:VOLTage<n>[:DC]:RANGe:PTPeak <NRf>
| MINimum | MAXimum
<n> [1] | 2 | 3 | 4
Note: Channel 3 not applicable for PM33x0B.
<NRf> The vertical sensitivity for channel <n> in peak-to-
peak volts, expressed in full scale (8 divisions).
MINimum Selects the minimum possible peak-to-peak value.
MAXimum Selects the maximum possible peak-to-peak value.
Query form:
SENSe:VOLTage<n>[:DC]:RANGe:PTPeak?
[MINimum | MAXimum]
Response:
<NR3>
<NR3> The vertical sensitivity for channel <n> in peak-to-
peak Volt, expressed in full scale (8 divisions).
If MINimum was specified, the minimum possible value is returned.
If MAXimum was specified, the maximum possible value is
returned.
Description:
Controls the vertical sensitivity for an input channel. The vertical sensitivity
(expressed in full scale, 8 divisions) for channel <n> is set to a value of <ptpeak>
volts. If a detectable probe is attached, the <ptpeak> value is considered to be at
the probe tip; otherwise the value is at the BNC plug.
The number of points with which a trace is written on the screen depends on the
resolution of the trace sample points (FORmat command). If the resolution is
8 bits, the number of points is 200 for the whole screen, which implies 200 / 8 = 25
points per division. If the resolution is 16 bits, the number of points is 200
*
256 =
51200 for the whole screen, which implies 51200 / 8 = 6400 points per division.