3 - 12 USING THE COMBISCOPE INSTRUMENTS
Examples:
MEASure:FALL:TIME? (@3)
Measures the time interval during which the pulse at channel 3 decreases
from 90% to 10% of its amplitude.
MEASure:RISE:TIME? 20,80
Measures the time interval during which the pulse at the default channel 1
increases from 20% to 80% of its amplitude.
The following measure functions and parameters can be programmed:
<measure_function><measure_parameters>
:AC
:AMPLitude
[:DC]
:FALL
:OVERshoot
:PREShoot
:TIME [<reference_low> [,<reference_high> [,<expected_time>
[,<time_resolution>]]]]
:FREQuency [<expected_frequency> [,<frequency_resolution>]]
:HIGH
:LOW
:MAXimum
:MINimum
:NDUTycycle <reference_middle>
:NWIDth <reference_middle>
:PDUTycycle <reference_middle>
:PERiod [<expected_period> [,<period_resolution>]]
:PTPeak
:PWIDth <reference_middle>
:TMAXimum
:TMINimum
:RISE
:OVERshoot
:PREShoot
:TIME [<reference_low> [,<reference_high> [,<expected_time>
[,<time_resolution>]]]]
Notes: - :DCYCle = alias for :PDUTycycle
- :FTIMe = alias for :FALL:TIME
- :RTIMe = alias for :RISE:TIME