COMMAND REFERENCE 4 - 121
After a
*
RST command, the trigger level is MAXimum and auto level peak-peak
is switched off.
Notice that there exists a coupling between programming the attenuator (vertical
sensitivity) and the trigger level. If the attenuator is changed, the trigger level is
also adapted to keep the signal display on the screen.
Programming tip:
First program the attenuator (SENSe:VOLTage:RANGe:PTPeak), and then the
trigger level (TRIGger:LEVel).
Example:
Send
*RST Resets the instrument.
Send
TRIGger:SOURce INTernal1 Trigger source becomes
channel 1.
Send
INITiate:CONTinuous ON Continuous initiation.
Send
SENSe:VOLTage:RANGe:PTPeak 8 1 V/div. sensitivity
Send
TRIGger:LEVel 0.2 Trigger level becomes 0.2 V.
Level peak-peak is also
switched off.
Send
TRIGger:LEVel:AUTO ON Switches level peak-peak on
and deactivates the trigger
level.
Front panel compliance:
The TRIGger:LEVel command is the remote equivalent of the front panel
TRIGGER LEVEL knob. The TRIGger:LEVel:AUTO command is the remote
equivalent of the front panel TRIGGER MAIN TB - level-pp on/off softkey menu.