Power Meter Operation
Configuring the Remote Interface
2-64 HP E4419B User’s Guide
Programming Language Selection
You can select one of two languages to program the power meter from the
remote interface. The language is SCPI when the power meter is shipped
from the factory. The other language is the HP 438A programming
language.
The power meter complies with the rules and regulations of the 1995.0
version of SCPI (Standard Commands for Programmable Instruments).
You can determine the SCPI version with which the power meter’s in
compliance by sending the SYSTem:VERSion? command from the remote
interface. You cannot query the SCPI version from the front panel.
The language selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
To select the interface language from the front panel:
1. Press , , .
2. Select the language you require from and .
To select the interface language from the remote interface use the:
SYSTem:LANGuage command