HP (Hewlett-Packard) HP E4419B Saw User Manual


 
HP4402.book : chapter2.fm 39 Fri Feb 12 15:32:45 1999
Power Meter Operation
Setting Offsets
HP E4419B User’s Guide 2-39
Setting Offsets
Setting Channel Offsets
The power meter can be configured to compensate for a signal loss or gain
in your test setup (for example, to compensate for the loss of a 10 dB
attenuator). This gain or loss, referred to as a channel offset, is added to
the measured power before the result is displayed.
Offsets are entered in dB. The allowable range of values is -100 dB to
+100 dB. A positive value compensates for a loss, and a negative value
compensates for a gain.
If either a channel or a display offset is set then “Ofs” is displayed.
To enter a channel offset which compensates for a loss or gain:
1. Press .
2. Press or depending on
the channel which you want to apply offsets to.
3. Press . The power meter displays the offset value in a pop
up window. Modify this offset (see below) as desired.
Use or to modify the digit on which the cursor is
currently positioned.
Use or to move to other digits.
4. To confirm your choice press .
5. is automatically set to “On” when a value is
entered using .
Channel offsets can be disabled and re-enabled simply by pressing
.
Setting Display Offsets
The display offset function provides a method for entering display offset
values.
If either a channel or a display offset is set then “Ofs” is displayed.
To enter a display offset on the currently selected window:
1. Press , .
System
Inputs
A Input Settings B Input Settings
Offset
dB
Offset Off On
Offset
Offset Off On
Rel
Offset
Offset