Menu Reference
Front Panel Menu Reference "Frequency/Cal Fac" Menu
3-14 HP E4419B User’s Guide
Press this softkey to enter the measurement calibration factor for
channel A in the range of 1% to 150%. The default value is 100%.
Use the , , and hardkeys to change the
calibration factor. To confirm your choice press . Entering a
calibration factor using this softkey is only possible when you are
using an HP 8480 series power sensor.
Press this softkey to enter the frequency of the signal to be
measured on channel B, in the range of 0.001 MHz to
999.999 GHz. The default value is 50 MHz. Use the , ,
and hardkeys to change the frequency. To confirm your
choice press the appropriate frequency unit. Entering a frequency
using this softkey is only possible when you have a sensor
calibration table selected or you are using an HP E-series power
sensor.
Press this softkey to enter the measurement calibration factor for
channel B in the range of 1% to 150%. The default value is 100%.
Use the , , and hardkeys to change the
calibration factor. To confirm your choice press . Entering a
calibration factor using this softkey is only possible when you are
using an HP 8480 series power sensor.
A Cal Fac
%
B Freq
B Cal Fac
%