HP (Hewlett-Packard) HP E4419B Saw User Manual


 
HP4402.book : chapter4.fm 12 Fri Feb 12 15:32:45 1999
Error Messages
Error Messages
4-12 HP E4419B User’s Guide
-330 Self-test Failed;Serial Interface Fault
Refer to “Serial Interface”, on page 2-77 if you require a description of
this test.
-350 Queue overflow
The error queue is full and another error has occurred which could not
be recorded.
-361 Parity error in program
The serial port receiver has detected a parity error and consequently,
data integrity cannot be guaranteed.
-362 Framing error in program
The serial port receiver has detected a framing error and consequently,
data integrity cannot be guaranteed.
-363 Input buffer overrun
The serial port receiver has been overrun and consequently, data has
been lost.
-410 Query INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous
data is not overwritten). The output buffer is cleared when power has
been off, or after *RST (reset) command has been executed.
-420 Query UNTERMINATED
The power meter was addressed to talk (that is, to send data over the
interface) but a command has not been received which sends data to the
output buffer. For example you may have executed a CONFigure
command (which does not generate data) and then attempted to read
data from the remote interface.
-430 Query DEADLOCKED
A command was received which generates too much data to fit in the
output buffer and the input buffer is also full. Command execution
continues but data is lost.
-440 Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a