Precision Trading 60 R3.0 Battery Charger User Manual


 
Operator Manual, Supercharger 60 R3.0
Page 27 of 62
6. OPERATING INSTRUCTIONS
6.1 GENERAL
6.1.1. Verify RESET
6.1.1.1. Verify that the reset button is depressed before turning power on.
6.1.2. Connect the Battery
6.1.2.1. Connect the battery cable to the batteries and the Temp-Plate.
6.1.2.2. If only one battery is used, connect the free plug to the shorted
receptacle provided in the Temp-Plate.
6.1.2.3. Observe polarity when using the single cell adaptor.
6.1.2.4. If battery polarity is reversed, a voltage fault condition will be
indicated and further operation of the instrument will be inhibited.
6.1.3. Verify Voltmeter
6.1.3.1. The digital voltmeter will read battery voltage if the selector switch is
in the internal position.
6.1.3.2. Select the scale consistent with the battery voltage.
6.1.4. Program the number of cells.
6.1.4.1. Enter the number of cells of the battery
6.1.4.2. When working with more than one battery enter the TOTAL number of
cells.
6.1.4.3. Note 1: In the full discharge mode the number of cells is ignored.
6.1.4.4. Note 2: In the Float and Peak voltage modes, the cell selector becomes
a nominal battery voltage selector. See xxx