Agilent 81130A Specifications
The following rules apply for pattern sequences:
The resolution of the segment length value depends on the frequency:
If the counted loop is used, the minimum length of the first segment is
twice the resolution.
An infinite loop over a single segment requires a minimum segment
length of twice the resolution.
Within a segment, PRBS is allowed to be combined with data type
PRBS, High or Low only.
PRBS must be combined with always the same data type in all
segments.
Human Interface
Overprogramming
Parameter values can be entered exceeding the specified range.
Warnings and Errors
Warning messages indicate potentially conflicting parameters due to
accuracy tolerances.
Error messages indicate conflicting parameters.
Resolution Frequency in MHz Period in ns
16 333.4 ... 666.7 1.500 ... 2.999
8 166.7 ... 333.3 3.000 ... 5.999
4 83.4 ... 166.6 6.000 ... 11.99
2 41.7 ... 83.3 12.00 ... 23.99
1 min. freq. ... 41.6 24.00 ... max. period