Agilent Technologies 660MHz Welding System User Manual


 
44
Programming Reference
SCPI Instrument Command List
Description
Use this command to select
STARTED
or
GATED
mode.
In the
gated mode
, the instrument triggers as long as the arming signal is
above (
:ARM:SENS POS
), or below (
:ARM:SENS NEG
) the selected
threshold level (
:ARM:LEV
).
In
started mode
, the instrument triggers on positive edge
(
:ARM:SENS POS
) or negative edge (
:ARM:SENS NEG
).
Command
:ARM:SENS
Long
:ARM[:SEQuence[1] | :STARt][:LAYer]:SENSe
Form
Set & Query
Parameter
POSitive | NEGative
*RST value
POS
Description
Use this command to select the edge or trigger level for the arming
signal.
The instrument triggers at the positive or negative cycle of the arming
signal.
Command
:ARM:SOUR
Long
:ARM[:SEQuence[1] | :STARt][:LAYer]:SOURce
Form
Set & Query
Parameter
IMMediate | EXTernal1 | MANual
*RST value
IMM
Description
Use this command to select the triggering mode of the instrument by
selecting the source of the arming signal:
Use
:ARM:MODE STARTed|GATed
to select the mode.
Triggering Source :ARM:SOURce Mode
Internal PLL
EXT INPUT
MAN key
IMMediate
EXTernal1
MANual
Continuous
Triggered | Gated by: EXT IN
Triggered | Gated by: MANKey