Programming Reference
SCPI Instrument Command List
Description
This command programs the amplitude current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute the
[:SOURce]:HOLD CURRent
command to enable the
[:SOURce]:CURRent
subsystem.
The available current range is limited by the specified voltage limits.
Example
:HOLD CURR
Enable CURRENT subsystem
:CURR1 75MA
Set OUTPUT 1 amplitude to 75 mA
Command
:CURR[1|2]:OFFSet
Long
[:SOURce]:CURRent[1|2][:LEVel][:IMMediate]:OFFSet
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
*RST value
0.0 µA (50 into 50 )
Specified Limits
3.8 V Outputs (50 into short): max. 152 mA typical
3.0 V Outputs (50 into short): max. 120 mA typical
Value coupling
Range coupling
Amplitude
Description
This command programs the offset current of the OUTPUT signal. Note
that to set the OUTPUT levels in terms of current, you first have to
execute the
[:SOURce]:HOLD CURRent
command to enable the
[:SOURce]:CURRent
subsystem
.
The available current range is limited by the specified voltage limits.
Example
:HOLD CURR
Enable CURRENT subsystem
:CURR1:OFF 50MA
Set OUTPUT 1 offset to 50 mA
Amplitude = High – Low
Offset =
High – Low
2