Agilent Technologies 660MHz Welding System User Manual


 
77
Programming Reference
SCPI Instrument Command List
Command
:VOLT[1|2]
Long
[:SOURce]:VOLTage[1|2][:LEVel][:IMMediate][:AMPLitude]
Form
Set & Query
Parameter
Numeric
Parameter suffix
V
with engineering prefixes.
*RST value
1.00 V
Specified Limits
Agilent 81131A: 0.10 Vpp to 3.80 Vpp
Agilent 81132A: 0.10 Vpp to 2.50 Vpp
Value coupling
Range coupling
Offset
Description
This command programs the amplitude voltage of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of voltage, you first have to
execute the
[:SOURce]:HOLD VOLTage
command to enable the
[:SOURce]:VOLTage
subsystem.
The available voltage range is limited by the specified current limits.
Example
:HOLD VOLT
Enable VOLTAGE subsystem
:VOLT1 2V
Set OUTPUT 1 amplitude to 2 V
Command
:VOLT[1|2]:OFFSet
Long
[:SOURce]:VOLTage[1|2][:LEVel][:IMMediate]:OFFSet
Form
Set & Query
Parameter
Numeric
Parameter suffix
V
with engineering prefixes.
*RST value
0.0 mV
High = Offset +
Amplitude
2
Low = Offset
2
Amplitude