Agilent Technologies 660MHz Welding System User Manual


 
99
Specifications
Agilent 81130A Specifications
Timing Specifications
The timing characteristics are measured at 50% amplitude at fastest
transitions in continuous mode and 50 load impedance.
NOTE
The Agilent 81130A is designed and recommended for an operation in the
frequency range of 170 kHz to 400/660 MHz. However it can be operated
in the extended range down to 1 kHz. Changes in specifications below
170 kHz are set in brackets [].
Period & Frequency
Period can also be entered as frequency.
Repeatability is typically four times better than accuracy.
Period & Frequency
Agilent 81130A with
Agilent 81131A
Agilent 81130A with
Agilent 81132A
Period range: 2.5 ns to 1 ms 1.5 ns to 1 ms
Frequency range: 1 kHz to 400.0 MHz 1 kHz to 660.0 MHz
Period/frequency
resolution:
4 digits, 2 ps best case
Period accuracy
a
:
a
In burst mode the first period may be decreased by 150 ps.
± 100 ppm [0.01%]
RMS-jitter:
(internal reference,
internal clock)
0.001% + 15 ps