Agilent Technologies 660MHz Welding System User Manual


 
68
Programming Reference
SCPI Instrument Command List
Specified limits
0.1  99.9%, depends on Width & Period.
Value coupling
Description
Use this command to program the dutycycle of the pulse signal. If you
want to set an absolute pulse-width use
[:SOURce]:PULSe:WIDTh[1|2]
.
If you want the pulse dutycycle to remain constant when the pulse period
is varied (rather than the absolute pulse width use)
[:SOURce]:PULSe:HOLD[1|2] DCYCle
Example
:PULS:DCYC1 25PCT
Set OUTPUT 1 dutycycle to 25%
:PULS:HOLD1 DCYC
Hold dutycycle constant with varying period
Command
:PULS:DEL[1|2]
Long
[:SOURce]:PULSe:DELay[1|2]
Form
Set & Query
Parameter
Numeric
Parameter suffix
S
with engineering prefixes. You can change the default unit using
[:SOURce]:PULSe:DELay[1|2]:UNIT
.
*RST value
0.0
Specified limits
0 to 3.00 µs
Value coupling
Width =
Duty Cycle
100
×
Period
Phase =
Delay
Period
×360
Delay% =
Delay
Period
×100