Agilent Technologies 8163A Work Light User Manual


 
Measurement Functions The SENSe Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 103
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
syntax: :SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
description: Returns the range setting for the module. For a return loss module, returns the power range
of the return loss diode.
parameters: none
response: The range setting as a float value in dBm
(110 value +30).
example: sens1:pow:rang? -2.00000000E+001<END>
affects: All power meters and return loss modules.
dual sensors: Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]
syntax: :SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]<wsp><value>[DBM]
description: Sets the power range for a retun loss module’s monitor diode.
The range changes at 10 dBm intervals. The corresponding ranges for linear measurements
(measurements in Watts) is given below:
Range
+30 dBm
+20 dBm
+10 dBm
0dBm
10 dBm
20 dBm
30 dBm
40 dBm
Upper Linear
Power Limit
1999.9 mW
199.99 mW
19.999 mW
1999.9 µW
199.99 µW
19.999 µW
1999.9 nW
199.99 nW
Range
50 dBm
60 dBm
70 dBm
80 dBm
90 dBm
100 dBm
110 dBm
Upper Linear
Power Limit
19.999 nW
1999.9 pW
199.99 pW
19.999 pW
1.999 pW
0.199 pW
0.019 pW
parameters: The range as a float value in dBm. The number is rounded to the closest multiple of 10, be-
cause the range changes at 10 dBm intervals. Units are in dBm.
response: none
example: sens1:pow:rang:mon -20DBM
affects: All return loss modules.
dual sensors: Master and slave channels are independent.