Agilent Technologies 8163A Work Light User Manual


 
GPIB Error Strings Error Codes
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 267
New -305
"Internal messaging error (StatCmdError)"
"Internal messaging error (StatCmdNotAllowed)"
"Internal messaging error (StatWrongLength)"
"Internal messaging error (StatWrongReceiver)"
"Internal messaging error (StatBufAllocError)"
"Internal messaging error (StatDPRamFull)"; }
"Internal messaging error (StatSemError)"
Description:
An error has occured in the instrument communication system. Please re-
port this error with a description of the circumstances that generated the
error and the configuration of the system.
New -306
"Channel doesn't support this command (StatCmdUnknownForSlave)"
Description:
Slave channels have limited functionality. The module supports this com-
mand, but the command must be sent to the master channel.
New -307
"Channel without head connection (StatHeadless)"
Description:
The channel supports this command, but it cannot be executed because
the optical measurement head is not plugged into the interface module.
Standard -310
“System error”
[Indicates that some error, termed “system error” by the device, has oc-
curred. This code is device-dependent.]
Standard -321
“Out of memory”
[An internal operation needed more memory than was available.]
New -322
"Flash programming error (StatFlashEraseFailed)"
"Flash programming error (StatFlashWriteFailed)"
"Flash programming error (StatFlashDataCntError)"
"Flash programming error (StatFlashDPAlgoFailed)"
Description:
An error has occured in a module. Please report this error with a descrip-
tion of the circumstances that generated the error and the configuration of
the system.
Table 18 Overview for Supported Strings
New/Old/Standard
Error
Number String