Agilent Technologies 8163A Work Light User Manual


 
Measurement Operations & Settings Signal Conditioning
158 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO:ALL
syntax: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO:ALL
description: Zero all available powermeter channels in the mainframe.
parameters: none
response: none
example: OUTP1:CORR:COLL:ZERO:ALL
affects: Powermeter modules; attenuator modules with power control, and return loss modules.
command:
:OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO?
syntax: :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO?
description: Queries the status of the last :OUTPutn[:CHANnel[m]]:CORRection:COLLection:ZERO opera-
tion.
parameters: none
response: integer 0 = OK, otherwise not OK.
example: OUTP:CORR:COLL:ZER0? 0<END>
affects: Attenuator modules with power control.