Agilent Technologies 8163A Work Light User Manual


 
Interface/Instrument Behaviour Settings The SYSTem Subsystem Instrument Setup and Status
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 77
command:
:SYSTem:HELP:HEADers?
syntax: :SYSTem:HELP:HEADers?
description: Returns a list of GPIB commands.
parameters: none
response: Returns a list of GPIB commands
example: syst:help:head? Returns a list of all GPIB commands
command:
:SYSTem:PRESet
syntax: :SYSTem:PRESet
description: Sets the mainframe and all installed modules to their standard settings. This command has
the same function as the Preset hardkey.
The following are not affected by this command:
•the GPIB (interface) state,
the backlight and contrast of the display,
the interface address,
the output and error queues,
the Service Request Enable register (SRE),
the Status Byte (STB),
the Standard Event Status Enable Mask (SESEM), and
the Standard Event Status Register (SESR).
parameters: none
response: none
example: SYST:PRES
command:
:SYSTem:TIME
syntax: :SYSTem:TIME<wsp><hour>,<minute>,<second>
description: Sets the instrument’s internal time.
parameters:
the first value is the hour (0 .. 23),
the second value is the minute, and
the third value is the seconds.
response: none
example: syst:time 20,15,30