Signal Conditioning Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 151
command:
:INPut[n][:CHANnel[m]]:WAVelength
syntax: :INPut[n][:CHANnel[m]]:WAVelength<wsp><value>[PM | NM | UM| MM | M] | MIN |
MAX | DEF
description: Sets the attenuator module’s operating wavelength.
This value is used to compensate for the wavelength dependence of the filter, and to calcu-
late a wavelegth dependent offset from the user offset table (if enabled).
parameters: <value>
MIN | MAX | DEF
The wavelength in meters (if you do not specify a unit).
Sets the wavelength to the module limits, or the module default.
response: none
example: INP1:WAV +1.55000000E-006
affects: All attenuator modules.
command:
:INPut[n][:CHANnel[m]]:WAVelength?
syntax: :INPut[n][:CHANnel[m]]:WAVelength?<wsp>MIN | MAX | DEF
description: Without the optional parameter, queries the operating wavelength of the attenuator.
parameters: MIN | MAX | DEF Queries the operating wavelength limits, or the module default.
response: 4 byte Intel floating point; wavelength in m.
example: INP1:WAV +1.55000000E-006<END>
affects: All attenuator modules.
command:
:OUTPutn[:CHANnel[m]]:APMode
syntax: :OUTPutn[:CHANnel[m]]:APMode<wsp><OFF(0) | ON(1)>
description: The use of this command is optional and has no effect on operation.
Included for compatibility with Agilent 8156A mainframe.
parameters: OFF or 0
ON or 1
response: none
example: OUTP1:APMode OFF
affects: All attenuator modules.