Agilent Technologies 8163A Work Light User Manual


 
Triggering - The TRIGger Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 183
trig:conf:ext #H2,#H0,#H0
trig2:outp dis
trig2:inp next
sour2:wav:swe star
trig1:outp stf
trig1:inp ign
sour1:wav:swe star
trig:conf:ext #H2,#H0,#H0 is described by Figure 4-1 and sets one bit:
for Node A Input Configuration:
Bit 1 - an event at slot 1 can trigger Node A. As trig1:outp stf is set, Node A
can be triggered if a sweep step finishes for a tunable laser module installed
in slot 1.
The following explanation explains the sequence with which actions are
triggered.
1 sour2:wav:swe star arms the sweep for for the tunable laser module in slot 2.
Because trig2:inp next is set, the module waits for a trigger until it performs the
first step of the sweep.
2
sour1:wav:swe star commands the tunable laser module in slot 1 to
start sweeping. Because trig1:inp ign is set, the module performs a
sweep as normal.
3 When the module in slot 1 finishes a step, because trig1:outp stf is set,
Node A is triggered.
4 Node A triggers all modules because the Output Matrix Configuration is
set to zero. Node A triggers the tunable laser module in slot 2 to perform
a sweep step because trig2:inp next is set.
5 The sequence starts again at step 3 and continues until the sweep ends.