Panasonic 35 Work Light User Manual


 
13 Troubleshooting
57
Symptoms Checks/Remedies
LED does not irradiate. Did you select the target channel for the product type? (When selected, the channel
switch is lit or flashing.)
Select the target channel when setting the product type.
If the target channel is selected but does not irradiate, check if the irradiation time
is set to 0.0sec for the channel program for the target product type.
If so, set the irradiation time to 0.0sec or longer.
When a warning message is displayed:
Temperature warning The channel connection indicator changes to orange (mixture of green and red)
when the LED head temperature has reached the warning level. If you continue the
operation in this condition, the product life will be significantly shortened, and the
temperature may exceed the upper limit, causing an emergency stop.
Set the UV intensity level (%) lower or improve the heat radiation performance
of the LED head to keep the temperature lower than the warning level.
Time warning When the total irradiation time of the LED reaches the point 30 hours before the
preset LED replacement point, the time warning is issued and x100hrs” will be
lit. When the time reaches the LED replacement point, the irradiation process will
make an emergency stop.
Prepare a new LED head. After replacement, clear the total irradiation time.
When an error occurs:
Interlock emergency stop
The interlock contact was opened
during irradiation.”
The emergency stop contact was opened during irradiation.
Check the emergency stop contact or the external input terminal block.
Close the contact and then restart irradiation.
* If the emergency stop contact is open, normal resetting is not available.
Hold down SET to forcibly reset the interlock.
Connection error
The LED was disconnected during
irradiation.”
It is possible that the connection cable or LED head was broken.
Reset the error status, check the connection of the cable and LED head with the
channel in question, and replace them if necessary.
LED wire broken
LED circuit opening was detected
during irradiation.”
The current detection was stopped during irradiation. It is possible that the
connection cable or LED head was broken.
Reset the error status, check the connection of the connection cable and LED
head with the channel in question, and replace them if necessary.
LED short-circuited
LED short-circuiting was detected
during irradiation.”
An overcurrent was detected during irradiation. It is possible that the connection
cable or LED head was short-circuited.
Reset the error status, check the connection of the cable and LED head with the
channel in question, and replace them if necessary.
LED temperature error
The LED temperature reached an
abnormal temperature.”
The LED temperature exceeded the upper limit.
Set the UV intensity level (%) lower or improve the heat radiation performance
of the LED head to keep the temperature lower than the warning level.
LED time error
The total irradiation time reached the
replacement point.”
The total irradiation time of the LED reached the preset LED replacement point.
(Or, it is possible that the LED replacement point was set ahead the total irradiation
time.)
Reset the error status, replace the LED head, and then clear the total irradiation
time. (Or, go to the mode setting screen and correctly set the LED replacement
point.)
LED calibration error
Calibration uncompleted”
Calibration was stopped before completion. Or, the calibration was not successful.
It is possible that the setting is too high, or there is a problem in the dedicated UV
sensor (option).
Reset the error status, check the connection of the UV sensor, review the setting,
and then retry calibration.